Tripter COMPACT Technical Specifications

SHOCK WAVE GENERATOR
Principle of Operation Underwater spark gap discharge
Reflector openings 181/145 mm
Spark gap Long-life electrodes
Water capacity 8 liters
Shock wave coupling Flexible membrane
SHOCK WAVE PARAMETERS
Voltage range Variable: 15-21 kV, 5-Stage
Energy High, Medium, Low
Pressure at focus 170-440 bar
Therapeutic focal zone 13x13x48 mm
Focal distance 135 mm
Shock wave triggering ECG or clock synchnorized
Weight 170 Kg
Floor space 100 x 62 cm
IMAGING
Fluoroscopy 9" mobile image intensifier, remote controlled
Zoom exposure 9", 6.5", 5"
ABC (Automatic Exposure Rate Control)
Fluoroscopic power 7mA, 110kVp
Operating modes Continuous or pulsed 0.1 to 4 sec.
Image Workstation Patient demographics comprehensive file
10 Frame disintegration process
In-put X-Ray or U.S.
Treatment image manipulations:
Zoom, rotate, smoothing, edge enhancement
Hundreds of images storage capacity (on-line)
Contrast enhancement, noise reduction
Ultrasound (Optional) Electronic convex sector scanning
3.5 MHZ Probe
Multi Point Dynamic Focusing
CONTROL UNIT
External monitoring ECG (Optional)
Indications Power, Water, R-ECG, Trigger
TREATMENT TABLE
Motorized 3 axis table Remote controlled
Urological table attachment
Floor space 100 x 72 cm
Table top (LxW) 190 x 64 cm
Vertical motion 27 cm, electrical
Horizontal movement 10 cm each, electrical
Weight 140 Kg
GENERAL
115/230VAC60/50Hz Operating temperature: 15o-35oC
Electrical Safety IEC 601.1, Class I, Type B
Electrical consumption 2000W
Minimum room size 3 x 4 meters (incl. patient transfer)


Tripter-COMPACT Regulatory Approval:
FDA - USA
MHW - Japan

Digiscope Regulatory Approval:
FDA - USA
CE - Europe
MHW- Japan

System configuration may vary due to local regulatory requirements